@@ -43,18 +43,18 @@ The following tables list the accuracy and precision specifications for Opentron
4343Flex 1-channel pipettes meet the following accuracy and precision specifications.
4444
4545<table >
46- <thead >
47- <tr>
48- <th>Pipette Type</th>
49- <th>Tip Size</th>
50- <th>Volume (µL)</th>
51- <th>Random Error (% CV)</th>
52- <th>Systematic Error (% D)</th>
53- </tr>
54- </thead >
46+ <thead >
47+ <tr >
48+ <th>Pipette Type</th>
49+ <th>Tip Size</th>
50+ <th>Volume (µL)</th>
51+ <th>Random Error<br>(% CV)</th>
52+ <th>Systematic Error<br>(% D)</th>
53+ </tr >
54+ </thead >
5555 <tbody >
5656 <tr>
57- <td rowspan="5"><b>1-Channel (1 –50 µL) </b></td>
57+ <td rowspan="5"><b>1–50 µL</b></td>
5858 <td>20 µL</td>
5959 <td>1</td>
6060 <td>3.50%</td>
@@ -85,7 +85,7 @@ Flex 1-channel pipettes meet the following accuracy and precision specifications
8585 <td>±1.25%</td>
8686 </tr>
8787 <tr>
88- <td rowspan="4"><b>1-Channel ( 5–1000 µL) </b></td>
88+ <td rowspan="4"><b>5–1000 µL</b></td>
8989 <td>50 µL</td>
9090 <td>5</td>
9191 <td>2.50%</td>
@@ -117,59 +117,71 @@ Flex 1-channel pipettes meet the following accuracy and precision specifications
117117Flex 8-channel pipettes meet the following accuracy and precision specifications.
118118
119119<table >
120- <thead >
120+ <thead >
121+ <tr >
122+ <th>Pipette Type</th>
123+ <th>Tip Size</th>
124+ <th>Volume (µL)</th>
125+ <th>Random Error<br>(% CV)</th>
126+ <th>Systematic Error<br>(% D)</th>
127+ </tr >
128+ </thead >
129+ <tbody >
121130 <tr>
122- <th>Pipette Capacity</th >
123- <th>Tip Capacity</th >
124- <th>Tested Volume</th >
125- <th>Accuracy %D</th >
126- <th>Precision %CV</th >
131+ <td rowspan="5"><b>1–50 µL</b></td >
132+ <td>20 µL</td >
133+ <td>1</td >
134+ <td>6.00%</td >
135+ <td>±10.00%</td >
127136 </tr>
128- </thead >
129- <tbody >
130137 <tr>
131- <td rowspan="3"><b>1–50 µL</b></td>
132138 <td>50 µL</td>
133- <td>1 µL</td>
134- <td>±10.00%</td>
139+ <td>1</td>
135140 <td>8.00%</td>
141+ <td>±10.00%</td>
136142 </tr>
137143 <tr>
138- <td>50 µL</td>
139- <td>10 µL</td>
140- <td>±2.50%</td>
144+ <td>20 µL</td>
145+ <td>5</td>
141146 <td>1.00%</td>
147+ <td>±2.50%</td>
142148 </tr>
143149 <tr>
144150 <td>50 µL</td>
151+ <td>10</td>
152+ <td>1.00%</td>
153+ <td>±2.50%</td>
154+ </tr>
155+ <tr>
145156 <td>50 µL</td>
146- <td>±1.25% </td>
157+ <td>50 </td>
147158 <td>0.60%</td>
159+ <td>±1.25%</td>
148160 </tr>
149161 <tr>
150162 <td rowspan="4"><b>5–1000 µL</b></td>
151163 <td>50 µL</td>
152- <td>5 µL</td>
164+ <td>5</td>
165+ <td>4.00%</td>
153166 <td>±8.00%</td>
154- <td>4%</td>
155167 </tr>
156168 <tr>
157169 <td>50 µL</td>
158- <td>50 µL</td>
159- <td>±2.50%</td>
170+ <td>50</td>
160171 <td>0.60%</td>
172+ <td>±2.50%</td>
161173 </tr>
162174 <tr>
163175 <td>200 µL</td>
164- <td>200 µL</td>
165- <td>±1.00%</td>
176+ <td>200</td>
166177 <td>0.25%</td>
178+ <td>±1.00%</td>
167179 </tr>
168180 <tr>
169181 <td>1000 µL</td>
170- <td>1000 µL</td>
171- <td>±0.70%</td>
182+ <td>1000</td>
172183 <td>0.15%</td>
184+ <td>±0.70%</td>
173185 </tr>
174186 </tbody >
175187</table >
@@ -179,66 +191,73 @@ Flex 8-channel pipettes meet the following accuracy and precision specifications
179191Flex 96-channel pipettes meet the following accuracy and precision specifications.
180192
181193<table >
182- <thead >
194+ <thead >
195+ <tr >
196+ <th>Pipette Type</th>
197+ <th>Tip Size</th>
198+ <th>Volume (µL)</th>
199+ <th>Random Error<br>(% CV)</th>
200+ <th>Systematic Error<br>(% D)</th>
201+ </tr >
202+ </thead >
203+ <tbody >
183204 <tr>
184- <th>Pipette Capacity</th >
185- <th>Tip Capacity</th >
186- <th>Tested Volume</th >
187- <th>Accuracy %D</th >
188- <th>Precision %CV</th >
205+ <td rowspan="5"><b>1–200 µL</b></td >
206+ <td>20 µL</td >
207+ <td>1</td >
208+ <td>3.00%</td >
209+ <td>±10.00%</td >
189210 </tr>
190- </thead >
191- <tbody >
192211 <tr>
193- <td rowspan="4"><b>1–200 µL</b></td>
194212 <td>50 µL</td>
195- <td>1 µL </td>
196- <td>±10 %</td>
197- <td>6 %</td>
213+ <td>1</td>
214+ <td>6.00 %</td>
215+ <td>±10.00 %</td>
198216 </tr>
199217 <tr>
200- <td>50 µL</td>
201- <td>5 µL </td>
202- <td>±4 %</td>
203- <td>2 %</td>
218+ <td>20 µL</td>
219+ <td>5</td>
220+ <td>1.00 %</td>
221+ <td>±4.00 %</td>
204222 </tr>
205223 <tr>
206224 <td>50 µL</td>
207- <td>50 µL </td>
208- <td>±1.5 %</td>
209- <td>1 %</td>
225+ <td>50</td>
226+ <td>1.00 %</td>
227+ <td>±1.50 %</td>
210228 </tr>
211229 <tr>
212230 <td>200 µL</td>
213- <td>200 µL </td>
214- <td>±1 %</td>
215- <td>1 %</td>
231+ <td>200</td>
232+ <td>1.00 %</td>
233+ <td>±1.00 %</td>
216234 </tr>
217235 <tr>
218236 <td rowspan="4"><b>5–1000 µL</b></td>
219237 <td>50 µL</td>
220- <td>5 µL </td>
221- <td>±10 %</td>
222- <td>5 %</td>
238+ <td>5</td>
239+ <td>5.00 %</td>
240+ <td>±10.00 %</td>
223241 </tr>
224242 <tr>
225243 <td>50 µL</td>
226- <td>50 µL</td>
227- <td>±2.5%</td>
244+ <td>50</td>
228245 <td>1.25%</td>
246+ <td>±2.50%</td>
229247 </tr>
230248 <tr>
231249 <td>200 µL</td>
232- <td>200 µL</td>
233- <td>±1.5%</td>
250+ <td>200</td>
234251 <td>1.25%</td>
252+ <td>±1.50%</td>
235253 </tr>
236254 <tr>
237255 <td>1000 µL</td>
238- <td>1000 µL </td>
239- <td>±1.5 %</td>
240- <td>1.5 %</td>
256+ <td>1000</td>
257+ <td>1.50 %</td>
258+ <td>±1.50 %</td>
241259 </tr>
260+ </tbody >
242261</table >
243262
244263## Pipette calibration
0 commit comments